EMGA-930 - Oxygen/Nitrogen/Hydrogen Combustion Analyzer

Overview

EMGA-930 O/N/H analyzer: It's a "New flagship model” for O/N/H analyzers and user friendly product.

This is a simultaneous oxygen, nitrogen & hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.

The oxygen is measured as carbon monoxide and carbon dioxide by two non-dispersive infrared detectors, the nitrogen by a thermal conductivity detector and the hydrogen with non dispersive Infra Red detector as H2O.

Principle and Gas flow diagram

The oxide in the sample reacts with the graphite crucible, and is extracted as carbon monoxide (CO), and carried together with the carrier gas. The extracted gas is directly analysed after the dust filter. According to the oxygen concentration, it is determined by CO or CO2 after pass throw oxidizer (Copper Oxide) with a non dispersive Infra-Red analyser (NDIR).

The Hydrogen is determined with a non dispersive Infra Red analyser as H2O after H2 passes through the oxidizer (Copper Oxide).

The Nitrogen in the sample is extracted as nitrogen gas (N2) and its concentration is determined by a thermal conductivity detector (TCD).

EMGA-930 Gas flow diagram

Features

Specifcation

EMGA-930: Oxygen/Nitrogen/Hydrogen analyzer

Principle
Oxygen: Non Dispersive Infrared detector 
Nitrogen: Thermal Conductivity detector(TCD)
Hydrogen: Non dispersive Infrared detector (as H2O)

Measurement range*: 
Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25%
*Up to 100% wt is possible by decreasing the sample weight.

Sample weight: 1.0 ± 0.1g

Sensitivity(Minimum reading): Oxygen/Nitrogen/Hydrogen: 0.001 ppm.

Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.02 ppm or RSD≦0.5% whichever is larger (Reference gas)
σn-1≦0.3 ppm or RSD≦1.0%  whichever is larger (Standard sample)
Hydrogen
σn-1≦0.04 ppm or RSD≦2% whichever is larger (Reference gas)
** Based on the sample weight 1.0 ± 0.1g and with automation functions described below

Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD

Power of furnace
Inert gas fusion in impulse furnace
Enable to set the generating power 0 – 8.0 kw

Sample introduction: Dual sample/flux introduction mechanism

Automation functions: Auto cleaner, Crucible loader

Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter

Sample ID: Enter up to 20 one-byte characters

Calibration

  1. One point or multi point calibration (Reference gas or standard samples)
  2. Calibration using previous analysis data
  3. Calibration curve correction function

Functions

  1. Display of real time extraction curve
  2. Analysis interruption
  3. Self diagnosis and alarm display
  4. Analysis of extraction curve (Superposition, difference etc analysis)
  5. Out put (RC-232C or TCP/IP)

Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.

Weight: 230kg
On carrying, it is divided in two units (Under 140kg/unit)

Computer: PC model operating OS Windows® XP(SP2) and beyond

Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz

Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)

Ground resistance: Less than 10Ω

Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band

Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit

Cooling mechanism: Radiator unit (Separation type)

Electronic balance (Option)
Enable connection with electronic balance with 1-0.01mg minimum display

Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg

Download Datasheet

EMGA-920 - Oxygen/Nitrogen Combustion Analyzer

 

Overview

EMGA-920 O/N analyzer: It's a "New flagship model” for O/N analyzers and user friendly product.

This is an oxygen and nitrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on.   This is a new generation model optimized to fit to user's requests.

The oxygen is measured as carbon monoxide and carbon dioxide by a non-dispersive infrared detector and the nitrogen by a thermal conductivity detector.

Features

Specification

EMGA-920:Oxygen/Nitrogen analyzer

Principle
Oxygen:Non Dispersive Infrared detector 

Nitrogen:Thermal Conductivity detector(TCD)

Measurement range*: 
Oxygen:0-5%, Nitrogen:0-3%

*Up to 100% wt is possible by decreasing the sample weight.

Sample weight: 1,0 ± 0,1g

Sensitivity(Minimum reading): Oxygen/Nitrogen: 0.0000001%(m/m)

Accuracy**(Repeatability)
Oxygen/Nitrogen

σn-1≦0.02%(m/m) or CV≦0.5% whichever is larger (Reference gas)
σn-1≦0.3%(m/m) or CV≦1.0%  whichever is larger (Standard sample)
** Based on the sample weight 1,0 ± 0,1g and with automation functions described below

Display
1) Measurement result: LCD or printout

2) Alarm message: LCD or printout
3) Flow sheet: LCD

Power of furnace
Iner gas fusion in impulse furnace

Enable to set the generating power 0 - 8,0kw

Sample introduction: Dual sample/flux introduction mechanism

Automation functions: Auto cleaner, Crucible loader

Integration conditions:
Preset time integration

Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter

Sample ID: Enter up to 20 one-byte characters

Calibration
1) One point or multi point calibration (Reference gas or standard samples)

2) Calibration using previous analysis data
3) Calibration curve correction function

Functions
1) Display of realtime extraction curve

2) Analysis interruption
3) Self diagnosis and alarm display
4) Analysis of extraction curve (Superposition, difference etc analysis)
5) Out put (RC-232C or TCP/IP)

Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.

Weight: 230kg
On carrying, it is devided in two units (Under 140kg/unit)

Computer: PC model operating OS Windows® XP(SP2) and beyond

Power
Main unit: AC200/220/230/240V ± 10%

Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz

Electric power consumption
Main unit:12kVA (MAX)

Auto cleaner: 1.5kVA (MAX)

Ground resistance: Less than 10Ω

Installation condition
Operation temperature: 5-40°C

Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, Linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band

Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa

Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe(O.D.6mm) and suitable connecter fitting within 5m of unit

Cooling mechanism: Radiator unit (Separation type)

Electronic balance (Option)
Enable connection with electronic balance with 1-0,01mg minimum display


Optional automatic voltage regulator (AVR)
Capacity:15kVA

Weight: 130kg

Schematics

EMGA-920 Dimensional Outline (mm)

Download Datasheet

EMGA-921 - Hydrogen Analyzer

 

Overview

EMGA-921 Hydrogen analyzer: It's a "New flagship model” for H analyzer and user friendly product.

This is an hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.

The hydrogen is measured as H2 by a thermal conductivity detector.

The Hydrogen in the sample is extracted as Hydrogen gas (H2) and its concentration is determined by a thermal conductivity detector (TCD).

EMGA-921 Gas flow diagram

Features

Specification

EMGA-921: Hydrogen analyzer

Principle
Hydrogen: Thermal Conductivity detector (TCD)as H2

Measurement range*
Hydrogen: 0-0.25%
*Possible to increase the range by reducing the sample weight.

Sample weight: 1.0 ± 0.1g

Sensitivity (Minimum reading): Hydrogen: 0.001 ppm.

Accuracy** (Repeatability)
Hydrogen
σn-1≦0.04 ppm or RSD≦0.5% whichever is larger (Reference gas)
** Based on the sample weight 1.0 ± 0.1g and with automation functions described below

Display

  1. Measurement result: LCD or printout
  2. Alarm message: LCD or printout
  3. Flow sheet: LCD

Power of furnace
Inert gas fusion in impulse furnace
Enable to set the generating power 0 – 8.0 kw

Sample introduction: Dual sample/flux introduction mechanism

Automation functions: Auto cleaner, Crucible loader

Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter

Sample ID: Enter up to 20 one-byte characters

Calibration

  1. One point or multi point calibration (Reference gas or standard samples)
  2. Calibration using previous analysis data
  3. Calibration curve correction function

Functions

  1. Display of real time extraction curve
  2. Analysis interruption
  3. Self diagnosis and alarm display
  4. Analysis of extraction curve (Superposition, difference etc analysis)
  5. Out put (RC-232C or TCP/IP)

Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.

Weight: 230kg
On carrying, it is divided in two units (Under 140kg/unit)

Computer: PC model operating OS Windows® XP(SP2) and beyond

Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz

Electric power consumption
Main unit: 12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)

Ground resistance: Less than 10Ω

Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band

Required gas
Ar carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit

Cooling mechanism: Radiator unit (Separation type)

Electronic balance (Option)
Enable connection with electronic balance with 1-0.01mg minimum display

Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg

Download Datasheet