EMGA-930 - Oxygen/Nitrogen/Hydrogen Combustion Analyzer
Overview
EMGA-930 O/N/H analyzer: It's a "New flagship model” for O/N/H analyzers and user friendly product.
This is a simultaneous oxygen, nitrogen & hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.
The oxygen is measured as carbon monoxide and carbon dioxide by two non-dispersive infrared detectors, the nitrogen by a thermal conductivity detector and the hydrogen with non dispersive Infra Red detector as H2O.
Principle and Gas flow diagram
The oxide in the sample reacts with the graphite crucible, and is extracted as carbon monoxide (CO), and carried together with the carrier gas. The extracted gas is directly analysed after the dust filter. According to the oxygen concentration, it is determined by CO or CO2 after pass throw oxidizer (Copper Oxide) with a non dispersive Infra-Red analyser (NDIR).
The Hydrogen is determined with a non dispersive Infra Red analyser as H2O after H2 passes through the oxidizer (Copper Oxide).
The Nitrogen in the sample is extracted as nitrogen gas (N2) and its concentration is determined by a thermal conductivity detector (TCD).
Features
- Super High Performance
- User-friendly software
- Fully Supported Accessories give simple operation
- Improved Mechanism
- Dual Sample/Flux introduction mechanism
Specifcation
EMGA-930: Oxygen/Nitrogen/Hydrogen analyzer
Principle
Oxygen: Non Dispersive Infrared detector
Nitrogen: Thermal Conductivity detector(TCD)
Hydrogen: Non dispersive Infrared detector (as H2O)
Measurement range*:
Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25%
*Up to 100% wt is possible by decreasing the sample weight.
Sample weight: 1.0 ± 0.1g
Sensitivity(Minimum reading): Oxygen/Nitrogen/Hydrogen: 0.001 ppm.
Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.02 ppm or RSD≦0.5% whichever is larger (Reference gas)
σn-1≦0.3 ppm or RSD≦1.0% whichever is larger (Standard sample)
Hydrogen
σn-1≦0.04 ppm or RSD≦2% whichever is larger (Reference gas)
** Based on the sample weight 1.0 ± 0.1g and with automation functions described below
Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD
Power of furnace
Inert gas fusion in impulse furnace
Enable to set the generating power 0 – 8.0 kw
Sample introduction: Dual sample/flux introduction mechanism
Automation functions: Auto cleaner, Crucible loader
Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter
Sample ID: Enter up to 20 one-byte characters
Calibration
- One point or multi point calibration (Reference gas or standard samples)
- Calibration using previous analysis data
- Calibration curve correction function
Functions
- Display of real time extraction curve
- Analysis interruption
- Self diagnosis and alarm display
- Analysis of extraction curve (Superposition, difference etc analysis)
- Out put (RC-232C or TCP/IP)
Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.
Weight: 230kg
On carrying, it is divided in two units (Under 140kg/unit)
Computer: PC model operating OS Windows® XP(SP2) and beyond
Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz
Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)
Ground resistance: Less than 10Ω
Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit
Cooling mechanism: Radiator unit (Separation type)
Electronic balance (Option)
Enable connection with electronic balance with 1-0.01mg minimum display
Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg
EMGA-920 - Oxygen/Nitrogen Combustion Analyzer
Overview
EMGA-920 O/N analyzer: It's a "New flagship model” for O/N analyzers and user friendly product.
This is an oxygen and nitrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.
The oxygen is measured as carbon monoxide and carbon dioxide by a non-dispersive infrared detector and the nitrogen by a thermal conductivity detector.
Features
- Super High Performance
- User-friendly software
- Fully Supported Accessories give simple operation
- Improved Mechanism
- Dual Sample/Flux introduction mechanism
Specification
EMGA-920:Oxygen/Nitrogen analyzer
Principle
Oxygen:Non Dispersive Infrared detector
Nitrogen:Thermal Conductivity detector(TCD)
Measurement range*:
Oxygen:0-5%, Nitrogen:0-3%
*Up to 100% wt is possible by decreasing the sample weight.
Sample weight: 1,0 ± 0,1g
Sensitivity(Minimum reading): Oxygen/Nitrogen: 0.0000001%(m/m)
Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.02%(m/m) or CV≦0.5% whichever is larger (Reference gas)
σn-1≦0.3%(m/m) or CV≦1.0% whichever is larger (Standard sample)
** Based on the sample weight 1,0 ± 0,1g and with automation functions described below
Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD
Power of furnace
Iner gas fusion in impulse furnace
Enable to set the generating power 0 - 8,0kw
Sample introduction: Dual sample/flux introduction mechanism
Automation functions: Auto cleaner, Crucible loader
Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter
Sample ID: Enter up to 20 one-byte characters
Calibration
1) One point or multi point calibration (Reference gas or standard samples)
2) Calibration using previous analysis data
3) Calibration curve correction function
Functions
1) Display of realtime extraction curve
2) Analysis interruption
3) Self diagnosis and alarm display
4) Analysis of extraction curve (Superposition, difference etc analysis)
5) Out put (RC-232C or TCP/IP)
Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.
Weight: 230kg
On carrying, it is devided in two units (Under 140kg/unit)
Computer: PC model operating OS Windows® XP(SP2) and beyond
Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz
Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)
Ground resistance: Less than 10Ω
Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, Linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe(O.D.6mm) and suitable connecter fitting within 5m of unit
Cooling mechanism: Radiator unit (Separation type)
Electronic balance (Option)
Enable connection with electronic balance with 1-0,01mg minimum display
Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg
EMGA-921 - Hydrogen Analyzer
Overview
EMGA-921 Hydrogen analyzer: It's a "New flagship model” for H analyzer and user friendly product.
This is an hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology's R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user's requests.
The hydrogen is measured as H2 by a thermal conductivity detector.
EMGA-921 Principle and Gas flow diagram
Features
- Super High Performance
- User-friendly software
- Fully Supported Accessories give simple operation
- Improved Mechanism
- Dual Sample/Flux introduction mechanism
Specification
EMGA-921: Hydrogen analyzer
Principle
Hydrogen: Thermal Conductivity detector (TCD)as H2
Measurement range*:
Hydrogen: 0-0.25%
*Possible to increase the range by reducing the sample weight.
Sample weight: 1.0 ± 0.1g
Sensitivity (Minimum reading): Hydrogen: 0.001 ppm.
Accuracy** (Repeatability)
Hydrogen
σn-1≦0.04 ppm or RSD≦0.5% whichever is larger (Reference gas)
** Based on the sample weight 1.0 ± 0.1g and with automation functions described below
Display
- Measurement result: LCD or printout
- Alarm message: LCD or printout
- Flow sheet: LCD
Power of furnace
Inert gas fusion in impulse furnace
Enable to set the generating power 0 – 8.0 kw
Sample introduction: Dual sample/flux introduction mechanism
Automation functions: Auto cleaner, Crucible loader
Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter
Sample ID: Enter up to 20 one-byte characters
Calibration
- One point or multi point calibration (Reference gas or standard samples)
- Calibration using previous analysis data
- Calibration curve correction function
Functions
- Display of real time extraction curve
- Analysis interruption
- Self diagnosis and alarm display
- Analysis of extraction curve (Superposition, difference etc analysis)
- Out put (RC-232C or TCP/IP)
Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.
Weight: 230kg
On carrying, it is divided in two units (Under 140kg/unit)
Computer: PC model operating OS Windows® XP(SP2) and beyond
Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz
Electric power consumption
Main unit: 12kVA (MAX)
Auto cleaner: 1.5kVA (MAX)
Ground resistance: Less than 10Ω
Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
Required gas
Ar carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe (O.D.6mm) and suitable connecter fitting within 5m of unit
Cooling mechanism: Radiator unit (Separation type)
Electronic balance (Option)
Enable connection with electronic balance with 1-0.01mg minimum display
Optional automatic voltage regulator (AVR)
Capacity:15kVA
Weight: 130kg