Ultima Expert

 

Over View

High resolution, high sensitivity and high stability ICP-OES spectrometer

  • The Ultima Expert offers the highest performance on the market and is specially designed to handle the most challenging applications.
  • Its high quality optical design which integrates a high density holographic grating and one meter focal length associated with the unique Total Plasma View feature offers the highest resolution of the market along with high sensitivity and stability for the most demanding applications.
  • Radial viewing and Total Plasma View providing lowest detection limits in ICP-OES spectrometry
  • Unique sheath gas feature for unrivaled stability on samples with High Total Dissolved Solids content, up to 30% dissolved solids
  • Continuous wavelength coverage from 120 to 800 nm with Far UV option for improved sensitivity for halogens analysis or alternative wavelengths
  • Unmatched spectral resolution minimizing interferences with <5 pm for wavelengths in the range 120-320 nm
  • Optional dual grating system, offering <6pm resolution up to 450 nm for line-rich matrices such as rare earth elements, geological or precious metals applications
  • Unique HiStab device to achieve exceptional stability for demanding applications
  • Patented HDD® detection providing sub-ppb to percent level in a single analysis
  • Dynamic range of up to 10 orders of magnitude using Image
  • Optional 0.5m or 1m polychromator for simultaneous analysis

Image Navigator

  • Image Navigator is designed for semi-quantitative analysis thanks to full spectrum acquisition.
  • Full spectrum display with multiple spectra display capability
  • Visual comparison with overlaid spectra display capability
  • Automated qualitative and semi-quantitative analysis

MASTER, find easily the right wavelengths

  • MASTER facilitates the method development by performing automatically selection of appropriate lines according to the concentration range of each element and the matrix. MASTER is based on the proprietary S3-base developed with real ICP spectra acquisitions and containing more than 50,000 wavelengths identified with full spectroscopic data.
    • Automatic wavelength selection
    • Synthetic spectra display
    • Sensitivity, dynamic range and interference level criteria to adjust line selection

S3-base viewer

  • S3-base viewer displays the entire S3-base database to help with identification of emission lines in unknown samples.
    • Full S3-base database available
    • Display of analyte wavelength and its relative intensity and detection limit
    • Display of the list of potential emission lines, including plasma inherent lines, in the vicinity of a selected wavelength with their relative intensities and detection limits
    • Graphic display of analyte emission line and potential emission lines for facilitated identification

Specifcation

Generator

Solid state, 40.68 MHz, water-cooled

Torch

Fully demountable torch, 3 mm internal diameter alumina injector, glass inner and outer tubes

Sample introduction

Concentric glass nebulizer and glass cyclonic spray chamber, 3 channels peristaltic pump, sheath gas device

Optical system

Thermally stabilized, 1 meter focal length, 2400 g/mm grating used in 1st and 2nd order with optical resolution <5 pm for 120-320 nm and <11 pm for 320-800 nm

Optional dual grating system

 with 4320 g/mm and 2400 g/mm used in the 1st order with optical resolution <6 pm for 120-450 nm and <11 pm for 450-800 nm.

Optional

0.5m or 1m polychromator

Detection

Patented HDD system

Download Datasheet

Ultima Expert LT

 

Overview

High performance affordable ICP-OES spectrometer

  • The Ultima Expert LT offers high performance at an affordable price. It can handle challenging applications due to its design.
  • Its high quality optical design which integrates a high density holographic grating and one meter focal length associated with the unique Total Plasma View feature offers high resolution along with high sensitivity.
  • Radial viewing and Total Plasma View providing lowest detection limits in ICP-OES spectrometry
  • Unique sheath gas feature for unrivaled stability on samples with High Total Dissolved Solids content, up to 30% dissolved solids
  • Continuous wavelength coverage from 120 to 800 nm with Far UV option for improved sensitivity for halogens analysis or alternative wavelengths
  • Unmatched spectral resolution minimizing interferences with <5 pm for wavelengths in the range 120-320 nm

S3-base viewer

  • S3-base viewer displays the entire proprietary S3-base, developed with real ICP spectra acquisitions and containing more than 50,000 wavelengths identified with full spectroscopic data, to help with identification of emission lines in unknown samples.
    • Full S3-base database available
    • Display of analyte wavelength and its relative intensity and detection limit
    • Display of the list of potential emission lines, including plasma inherent lines, in the vicinity of a selected wavelength with their relative intensities and detection limits
    • Graphic display of analyte emission line and potential emission lines for facilitated identification

Specification

Generator

Solid state, 40.68 MHz, water-cooled

Torch

Fully demountable torch, 3 mm internal diameter alumina injector, glass inner and outer tubes

Sample introduction

Concentric glass nebulizer and glass cyclonic spray chamber, 3 channels peristaltic pump, sheath gas device

Optical system

Thermally stabilized, 1 meter focal length, 2400 g/mm grating used in 1st and 2nd order with optical resolution <5 pm for 120-320 nm and <11 pm for 320-800 nm

Detection

Dual PMT detection

Download Datasheet

ICP Neo

 

Overview

  • ICP Neo software for HORIBA Scientific ICP-OES spectrometers is designed to facilitate method development, samples measurements and results management.
  • ICP Neo is compliant with new Microsoft platforms and delivers powerful tools for samples measurement with new HDD mode integrated for standard measurement, advanced  Quality Control Protocols and retrospective analysis with respect to the integrity of raw results to match with good laboratory practices requirements.

Specification

  • Available for all Ultima family ICP-OES spectrometers.
  • CPU05 electronics is required. All other HORIBA Scientific ICP-OES an electronic upgrade is available.
  • Method development
    • Interference free lines by visual display using S3 wavelengths database
    • Unique HDD mode to remove the need for optimization of High Voltage
    • Multiple measurement parameters with Gaussian, Maximum and Mean mode
    • Multiple calibration mode
  • Sequence
    • SmartRinse for automatic monitoring of rinse efficiency between 2 samples
    • Automatic control of the quality with limits on correlation coefficient and on recalculated concentrations
    • Use of multiple methods in a single sequence
  • Results
    • Archive capability
    • Reprocess capability with integrity of raw results
    • Reports with user’s selectable data
    • Customization of reports with logo, instrument name…
    • Print and export results for LIMS
  • Quality Control Protocols
    • Fully US EPA compliant QC protocols with ICV, CCV, LCS, Interference Check, Paired samples, Spikes, and more.
    • Automated actions in case of value out of specifications from stop analysis to calibrate and repeat samples from the last valid QC
  • Ease of use
    • Multiple users capability
    • Contextual tool bar adapting the content with the window selected
    • User defined windows display on the screen with possibility to hide windows
    • User defined maintenance warnings
    • Diagnostics
      • Real-time information on instrument status with filter selection, slits positions, grating
      • position and PMT selection
      • Zero order and Reference acquisition with graphic display

Download Datasheet